In their recent article, Kreutzer et al.1 evaluated intraocular pressure (IOP) during standard and microcoaxial phacoemulsification. I first used intraocular manometry during phacoemulsification in 1985 and found, as did Kreutzer et al., that IOP is frequently very elevated for short periods of time. A related observation is that during vitreoretinal procedures, it is not uncommon for complete blanching of the optic nerve with closure of retinal arterioles to occur and to persist for significantly longer periods of time than during phacoemulsification procedures. Kreutzer et al. state that they believe monitoring IOP can prevent the surgeon from inducing excessive IOP elevation and thereby prevent damage to eyes with compromised retinal perfusion or previous pathology of the nerve fiber layer. For decades, other surgeons have also expressed concerns about damage from elevated IOP. However, there is no evidence to indicate that such damage occurs during phacoemulsification. For more than 30 years, I have routinely performed the procedure with an infusion bottle height of 130 to 140 cm with virtually no leakage from the phaco or side-port incision; the problem has been nonexistent in my experience. There must, of course, be some duration of IOP elevation that would induce optic nerve and/or retinal damage, but that parameter is unknown. It is obviously much longer than the time of IOP elevation during the procedures in question.