Key result
Video capsule endoscopy shows a ~0% adverse cardiac event or device malfunction rate in CIEDs.
Why the study?
Video capsule endoscopy remains contraindicated by the FDA and main device manufacturers in patients with cardiac implantable electronic devices due to theoretical electromagnetic interference and possible device malfunction.
Is video capsule endoscopy safe and effective in patients with implantable cardiac devices (pacemakers, defibrillators, and left ventricular assist devices)?
Meta-Analysis (n=735)
Is video capsule endoscopy safe and effective in patients with implantable cardiac devices (pacemakers, defibrillators, and left ventricular assist devices)?
Video capsule endoscopy is safe and maintains an 85% completion rate and 70% diagnostic yield without causing cardiac device malfunction or adverse cardiac events in patients with PPMs, ICDs, or LVADs, supporting its use despite historical FDA and manufacturer contraindications.
Answers are grounded in this paper and cite their sources.
Supports clinical use of video capsule endoscopy in device patients; challenges longstanding contraindications and extends the safety evidence base.
Background: To date, video capsule endoscopy (VCE) is still contraindicated by the FDA and the main manufacturers of Cardiac Implantable Electronic Devices (CIED) in patients with CIED, given a theoretical electromagnetic interference and possible device malfunction. Objectives: The objective of this study was to assess the safety profile and efficacy of VCE in patients with implantable cardiac devices through analyzing the risk of mutual interference. Methods: A systematic review of PubMed, Web of Science, and Embase databases was conducted. Peer-reviewed original articles, published in the English language and containing "capsule endoscopy" AND "pacemaker", "defibrillator" OR "left ventricular assist device" as keywords, were selected. Studies performed in vitro, isolated case reports, and abstracts/posters were excluded. Results: A total of 735 VCE procedures were performed in patients with cardiac devices in various clinical settings. Cardiac events were not seen in any case. Interference on capsule images transmission was noted in 5 cases (left ventricular assist device (LVAD)) where few images were lost when the capsule was closest to the device. Finally, interference between capsule and telemetry leads was noted in 6 cases (4 Permanent Pacemakers (PPM), 2 Implantable Cardioverter-Defibrillator (ICD)) leading to image artifacts. Discussion: Adverse cardiac events were not seen in any study. Loss of images occurred when the VCE was in proximity to the device (only with LVAD) or after telemetry leads installation without affecting the completion rate and diagnostic yield of VCE. Conclusion: VCE is safe and remains efficient in patients with cardiac devices. If cardiac monitoring is required, wired systems are preferable.
No takes yet. Share an insight, caveat, or question.
Tabet et al. (2019) conducted a meta-analysis in Patients with implantable cardiac devices (PPM, ICD, LVAD) undergoing video capsule endoscopy (n=735). Video capsule endoscopy (VCE) was evaluated on Safety of the VCE procedure defined as successful completion without cardiac adverse events or cardiac device malfunction. Video capsule endoscopy in patients with implantable cardiac devices had a 0% rate of adverse cardiac events or device malfunction across 735 pooled procedures.