Key result
ViV-TAVR achieves 100% procedural success without mortality in high-risk degenerated stentless aortic bioroots.
Why the study?
Open surgical repair of a failed VSARR or stentless bio-ARR entails significant operative risks, and whether ViV-TAVR is feasible in these patients remains unclear due to concerns regarding an ill-defined aortic annulus and coronary obstruction.
Is valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) feasible and safe in high-risk patients with a degenerated stentless bioroot or previous VSARR?
Observational (n=12)
No
Is valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) feasible and safe in high-risk patients with a degenerated stentless bioroot or previous VSARR?
ViV-TAVR is a clinically effective and feasible option with favorable short- and intermediate-term outcomes for high-risk patients with failed stentless bio-ARR or previous VSARR.
BACKGROUND: Open surgical repair of a failed valve-sparing aortic root replacement (VSARR) or stentless bioroot aortic root replacement (bio-ARR) entails significant operative risks. Whether valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) is feasible in patients with a previous VSARR or stentless bio-ARR remains unclear, given lingering concerns about the ill-defined aortic annulus in these patients and the potential for coronary obstruction. We present our experience with patients who had a previous VSARR or stentless bio-ARR and underwent ViV-TAVR to repair a degenerated aortic valve with combined valvular disease, aortic insufficiency and aortic stenosis. METHODS: In this retrospective data review, we identified and analyzed consecutive patients with a previous VSARR or stentless bio-ARR who underwent ViV-TAVR between December 1, 2014 and August 31, 2019. RESULTS: ViV-TAVR was performed in twelve high-risk patients with previous VSARR or bio-ARR during the study period. Of these, seven received Medtronic Freestyle porcine stentless bioprosthetic aortic roots, three received homograft aortic roots, one underwent a Ross procedure and one underwent VSARR. ViV-TAVR restored satisfactory valve function in all patients, and technical success was 100%. No patient had more than mild regurgitation after implantation. No thirty-day mortality was seen. One patient had major bleeding after transapical access, one patient had a transient ischemic stroke, and one patient needed permanent pacemaker implantation. At a median last follow-up of 21.5 months (interquartile range, 9.0-69.0 months), all patients remained alive and had satisfactory valve function. CONCLUSIONS: In this study, ViV-TAVR was a clinically effective option for treating patients with a failed stentless bio-ARR or previous VSARR. Short-term and intermediate-term results after these procedures were favorable. These findings may have important implications for treating high-risk patients with structural aortic root deterioration and call for better transcatheter heart valves that are suitable for treating aortic insufficiency.
No takes yet. Share an insight, caveat, or question.
Çekmecelioğlu et al. (2021) conducted an observational in Degenerated stentless aortic bioroots (n=12). Valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) was evaluated on All-cause mortality at 1 year. Valve-in-valve transcatheter aortic valve replacement achieved 100% procedural success and 0% mortality at a median follow-up of 21.5 months in high-risk patients with degenerated stentless aortic bioroots.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: