RATIONALE AND OBJECTIVES: The objective of this study was to investigate the potential for artifact-free coronary magnetic resonance angiography (cMRA) in the presence of dedicated metallic MR stents in vitro and in a swine model. METHODS: All investigations were performed at 1.5 T, applying a standard cMRA gradient echo sequence with a T2 preparation pulse. Two prototypes of each hand-woven, mechanically woven, and lasered Aachen Resonance Coronary MR Stents made out of an MR-compatible metallic alloy and dilated to 2.5 mm and 4 mm were examined in a water bath. RESULTS: Artifact behavior was judged independently by 2 radiologists as showing "no artifacts" for all tested stent types. Signal-to-noise ratios inside and outside of the stents were measured yielding a Pearson correlation coefficient of 0.98 (y = 1.22 + 0.92x). Nineteen stents (8 hand woven, 3 mechanically woven, 8 lasered) were deployed in coronary arteries of 19 domestic pigs and were examined by cMRA. Artifact behavior of the stents was analyzed by measuring the signal-to-noise ratio at the stent positions and compared with signal-to-noise ratio measurements outside of the stents, yielding a Pearson correlation coefficient of 0.90 (y = -0.75 + 1.06x). CONCLUSIONS: All 3 prototypes of coronary MR stents allowed complete visualization of the stent lumen and consequently determination of stent patency by cMRA.
No takes yet. Share an insight, caveat, or question.
Buecker et al. (2004) studied this question.