Key result
A novel 'J' shaped 6F bipolar electrode catheter allowed for rapid percutaneous insertion and stable temporary atrial pacing without fluoroscopy in 15 patients.
Why the study?
Does a novel 'J' shaped 6F bipolar electrode catheter allow for rapid and stable temporary atrial pacing without fluoroscopy?
Does a novel 'J' shaped 6F bipolar electrode catheter allow for rapid and stable temporary atrial pacing without fluoroscopy?
A novel 'J' shaped 6F bipolar electrode catheter allows for rapid, stable temporary atrial pacing via the subclavian vein without the need for fluoroscopy.
Answers are grounded in this paper and cite their sources.
May support fluoroless temporary atrial pacing in select cases; hypothesis-generating and requires randomized validation before practice change.
A permanently formed "J" shaped 6F bipolar electrode catheter specially designed to facilitate temporary atrial pacing was constructed and tested in 15 patients. A pair of rigid, "orienting wings" was fixed to the catheter 28 cm from a large spherical tip electrode to establish directional orientation of the "J" while in the right atrium. Rapid percutaneous introduction is possible with a peel-away sheath via the subclavian vein. Positioning in the right atrium can be accomplished without fluoroscopy. The catheter could be inserted and positioned against the right atrial wall generally in less than three minutes and required little or no additional manipulation for capture. No instance of loss of capture was documented during subsequent continuous monitoring as the catheter remained in place from one to thirteen days. Ease of insertion of this electrode catheter, as well as stability, provided an effective method for temporary atrial pacing without fluoroscopy. This novel catheter design warrants testing in a large patient population by others and could have a potential application to other types of catheter needs.
No takes yet. Share an insight, caveat, or question.
Sloman et al. (1981) studied Requirement for temporary atrial pacing (n=15). J shaped 6F bipolar electrode catheter was evaluated on Successful insertion, positioning, and maintenance of capture without fluoroscopy. A novel 'J' shaped 6F bipolar electrode catheter allowed for rapid percutaneous insertion and stable temporary atrial pacing without fluoroscopy in 15 patients.
Synapse has enriched 10 related papers. Consider them for comparative context: