Key result
Percutaneous tricuspid valve replacement achieved a 98% procedural success rate, with 100% of patients having tricuspid regurgitation ≤ 2+ at discharge and a 30-day mortality of 2.2%.
Why the study?
Transcatheter tricuspid valve replacement is a novel technique for symptomatic tricuspid regurgitation, prompting an evaluation of initial experience and short-term outcomes in Spain.
Does transcatheter tricuspid valve replacement improve tricuspid regurgitation and clinical outcomes in patients with symptomatic tricuspid regurgitation?
Observational (n=48)
Yes
Does transcatheter tricuspid valve replacement improve tricuspid regurgitation and clinical outcomes in patients with symptomatic tricuspid regurgitation?
Transcatheter tricuspid valve replacement demonstrates high procedural success, significant short-term reduction in tricuspid regurgitation, and low 30-day mortality in a real-world Spanish registry.
Answers are grounded in this paper and cite their sources.
Supports procedural feasibility in tricuspid regurgitation; leaves open durability, safety, and comparative effectiveness in randomized trials.
BACKGROUND: Transcatheter tricuspid valve replacement (TTVR) is a novel technique for the treatment of symptomatic tricuspid regurgitation (TR). AIMS: This study aims to present the initial experience in Spain and report short-term outcomes of this therapy. METHODS: This was a multicenter, retrospective registry including 48 patients who underwent percutaneous tricuspid valve replacement between April 2022 and September 2024. The primary efficacy objectives were procedural success and the degree of regurgitation at discharge. The primary safety objectives included all-cause mortality, hospitalization for heart failure, and other major adverse cardiovascular events. RESULTS: A total of 48 patients were included (79 [72-82] years, 75% women, TRISCORE 5 [4-6] points). Sixty-three percent of the patients were in New York Heart Association (NYHA) functional class ≥ III. TR was predominantly of secondary etiology (87%). Procedural success was 98%. The entire treated population achieved TR ≤ 2+ at discharge (100%), and TR ≤ 1+ was achieved in 94% of patients. A permanent pacemaker was implanted in 15.8% of patients. At 30 days, overall mortality was 2.2%, 82% of patients were in NYHA class I/II, and 100% and 94% had TR grades ≤ 2+ and ≤ 1+, respectively. CONCLUSIONS: Percutaneous tricuspid valve replacement proved to be effective and safe, with a high procedural success rate and a low incidence of adverse events. At 30 days, a significant reduction in TR grade was observed, along with an improvement in functional class.
No takes yet. Share an insight, caveat, or question.
Echarte‐Morales et al. (2025) conducted an observational in Symptomatic tricuspid regurgitation (n=48). Percutaneous tricuspid valve replacement was evaluated on Procedural success. Percutaneous tricuspid valve replacement achieved a 98% procedural success rate, with 100% of patients having tricuspid regurgitation ≤ 2+ at discharge and a 30-day mortality of 2.2%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: