Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
January 1, 2018Northern Clinics of IstanbulOpen Access

Validation of a smartphone based electrocardiography in the screening of QT intervals in children

View Full Paper
Ask AI
Bookmark
Share

Key result

AliveCor smartphone electrocardiography recordings showed a significant correlation with standard 12-lead electrocardiography for measuring corrected QT intervals in children (r=0.57, p<0.001).

Why the study?

Does a smartphone-based electrocardiography device accurately measure QT and corrected QT intervals compared to a standard 12-lead electrocardiograph in children?

Population

285 children presenting to the clinic, mean age 9.8±4.9 years

Comparison

10-s smartphone electrocardiography tracing from… vs Standard 12-lead electrocardiograph obtained…

Design

Cross-sectional, Measurements of QT and corrected QT intervals performed in a…

Authors

MKMehmet Karacan

Discussion

Loading...

Member takes

Overview

May support smartphone ECG for pediatric QTc screening where standard ECG is unavailable; leaves open validation for clinical decisions.

Study Design

Type

Observational (n=285)

Blinding

Blinded evaluators

Multicenter

No

Structured PICO

Does a smartphone-based electrocardiography device accurately measure QT and corrected QT intervals compared to a standard 12-lead electrocardiograph in children?

P
Population
285 pediatric and adolescent patients (mean age 9.8 years) presenting to a pediatric cardiology clinic underwent simultaneous 12-lead and smartphone-based electrocardiography to assess QT intervals.
I
Intervention
10-s smartphone electrocardiography tracing from a smartphone-based device (AliveCor)
C
Comparator
Standard 12-lead electrocardiograph obtained simultaneously
O
Outcome
Measurement and correlation of QT and corrected QT (QTc) intervals using Bazett's formulasurrogate

Main Result

Effect estimate: Pearson's correlation coefficient: 0.57

Absolute Event Rate: 415% vs 419%

p-value: p=<0.001

Smartphone-based electrocardiography provides QT and QTc measurements that correlate significantly with standard 12-lead ECGs, suggesting it is a viable tool for screening congenital long QT syndrome in children.

Limitations

  • 7.3% of transmitted tracings were not interpretable due to noise, motion artifact, or unclear T-wave termination.
  • Corrected QT intervals measured with Bazett's formula at variable heart rates have low sensitivity and accuracy, which may have influenced the discrepancy in results.
  • Lack of technicians' initial assessment of electrocardiography led to noisy chest recordings that were not repeated.
  • Larger population studies and cost-effectiveness analyses are needed before establishing it as a practical screening tool.

Cite This Study

Mehmet Karacan (2018) conducted an observational in Long QT syndrome screening (n=285). Smartphone-based electrocardiography (AliveCor) vs. Standard 12-lead electrocardiography was evaluated on Correlation of corrected QT intervals between 12-lead electrocardiographs and AliveCor recordings (Pearson's correlation coefficient: 0.57, p=<0.001). AliveCor smartphone electrocardiography recordings showed a significant correlation with standard 12-lead electrocardiography for measuring corrected QT intervals in children (r=0.57, p<0.001).

synapsesocial.com/papers/6aa4a2ec997531f8901426aahttps://doi.org/10.14744/nci.2018.44452
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1SPEAR Trial: Smartphone Pediatric ElectrocARdiogram Trial2015 · 75 citations
  2. 2Electrocardiogram Screening for Disorders That Cause Sudden Cardiac Death in Asymptomatic Children: A Meta-analysis2012 · 87 citations
  3. 3Cost effectiveness of neonatal ECG screening for the long QT syndrome2006 · 117 citations
  4. 4Wireless Smartphone ECG Enables Large‐Scale Screening in Diverse Populations2015 · 222 citations
  5. 5Heart Rate-Dependence of QTc Intervals Assessed by Different Correction Methods in Patients with Normal or Prolonged Repolarization2009 · 52 citations