Key result
Multi-channel lossless ECG compression system achieves ~69μW power consumption.
Why the study?
Long-term ECG monitoring produces large amounts of data, creating a need for effective lossless ECG compression systems to reduce storage space.
Answers are grounded in this paper and cite their sources.
A novel hardware architecture for multi-channel lossless ECG compression demonstrates low power consumption and hardware complexity, suitable for long-term ECG monitoring.
Electrocardiogram (ECG) is used to record the electrical activity of heart. If the instrument monitoring the signal for a long time, it will produce large amount of data. So, the effective lossless ECG compression system can help to reduce the storage space. This brief presents a hardware architecture of multi-channel lossless ECG compression system. The system is based on the algorithm including multi-channel linear prediction and adaptive linear prediction. Also, Golomb rice code (GRC) is used for entropy coding. The hardware implementation has been designed with low hardware complexity usage while using optimum hardware resources. And the architecture can process multiple channels in parallel that can obtain the high throughput. PTB database has been used for verification and testing purposes. This design was implemented in TSMC 180nm. The implementation results show the gate count is 476K and power consumption is 69.18μW while working frequency is 1 KHz.
No takes yet. Share an insight, caveat, or question.
Tsai et al. (2021) studied this question. Hardware architecture of multi-channel lossless ECG compression system was evaluated on Gate count, power consumption, and working frequency. The proposed multi-channel lossless ECG compression system implemented in TSMC 180nm achieved a gate count of 476K and power consumption of 69.18μW at a working frequency of 1 KHz.
Synapse has enriched 10 related papers. Consider them for comparative context: