Key result
Cross-talk manifesting as inappropriate ventricular inhibition occurred in 11.3% of patients with dual anodal ring Gemini 415A pacemakers compared to 0% with single anodal ring models.
Why the study?
Does the design of the Gemini 415A DDD pacemaker (dual vs. single anodal rings) affect the incidence of cross-talk?
Observational (n=89)
Does the design of the Gemini 415A DDD pacemaker (dual vs. single anodal rings) affect the incidence of cross-talk?
Cross-talk in the Gemini 415A DDD pacemaker is a design-dependent phenomenon occurring only in models with dual interconnected anodal rings when used with non-Cordis leads.
Answers are grounded in this paper and cite their sources.
Dual anodal ring design may increase cross-talk risk with incompatible leads; leaves open whether modern pacemakers share similar vulnerabilities.
A unique mechanism of "cross-talk" has recently been described in the Gemini 415A DDD pacemaker in which cross-inhibition of the ventricular output stimulus may occur in the early Gemini 415A models which were manufactured with dual anodal rings interconnected by a wire. The Gemini 415A units with dual anodal rings were implanted under investigational protocol in 53 patients from September 1982 to August 1984. To date, cross-talk has been discovered in six patients during clinical follow-up, manifesting as inappropriate ventricular inhibition. Syncope was documented in one patient. Non-Cordis atrial and ventricular leads were present in all six patients and five patients had at least one chronic lead. Cross-talk was neither evident nor provokable in Gemini 415A units which had compatible Cordis lead systems. Reduction of the atrial output and/or the ventricular sensitivity prevented cross-inhibition in three patients. Cross-talk persisted in three patients despite programming to the lowest atrial output and/or ventricular sensitivity parameters. Cross-talk, spontaneous or provoked, has not occurred in 36 patients having currently designed 415A pulse generators manufactured with a single anodal ring in the ventricular connector channel. We conclude that: this unique form of cross-talk is "design-dependent" and occurs only in Gemini 415As with dual interconnected anodal rings and cannot occur in 415As manufactured with a single anodal ring; this form of cross-talk has occurred only in 415As with non-Cordis leads and appears to be a time-dependent development; reduction of the atrial output or use of pacing modes without atrial pacing or ventricular sensing can prevent cross-talk.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
BEAVER et al. (1986) conducted an observational in Pacemaker implantation (n=89). Gemini 415A DDD pacemaker with dual anodal rings vs. Gemini 415A DDD pacemaker with single anodal ring was evaluated on Cross-talk (inappropriate ventricular inhibition). Cross-talk manifesting as inappropriate ventricular inhibition occurred in 11.3% of patients with dual anodal ring Gemini 415A pacemakers compared to 0% with single anodal ring models.