Does activation of the atrial sense marker function cause crosstalk in patients with Intermedics Cosmos DDD pulse generators?
Does activation of the atrial sense marker function cause crosstalk in patients with Intermedics Cosmos DDD pulse generators?
Crosstalk due to atrial sense markers in DDD pacemakers can be mitigated by reprogramming ventricular sensitivity or atrial output.
Answers are grounded in this paper and cite their sources.
Alerts to crosstalk risk in legacy Intermedics Cosmos devices amenable to reprogramming; case report leaves open applicability to modern pacemakers.
We studied the occurrence and characteristics of crosstalk related to the atrial sense marker function of the Intermedics Cosmos DDD pulse generator in 29 patients. Upon activation of the atrial sense markers, the pulse generator delivers a series of markers in the form of triggered atrial stimuli at 0.025 ms in duration at the programmed voltage output of the atrial channel. Under certain circumstances, these atrial sense marker stimuli may cause crosstalk when they are sensed by the ventricular sensing amplifier. This form of crosstalk may be eliminated in most cases by decreasing ventricular sensitivity and/or atrial output voltage.