Key result
An implanted atrial bipolar automatic burst tachycardia-terminating pulse generator caused inappropriate burst stimulation due to double sensing of P and QRS signals during normal sinus rhythm.
Case Report (n=2)
Far-field QRS signals can lead to double sensing and inappropriate burst stimulation in patients with atrial automatic tachycardia-terminating pulse generators.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to double-sensing risks in atrial burst devices; leaves open need for refined sensing algorithms.
This report describes two cases of double sensing of the P and QRS signals by an implanted atrial bipolar automatic burst tachycardia-terminating pulse generator (Intermedics CyberTach 60) used for the treatment of supraventricular tachycardia. Double sensing occurred during normal sinus rhythm at a rate slower than the tachycardia detection criterion and caused inappropriate delivery of burst stimulation. These observations underscore the importance of the far-field QRS signal in the detection of supraventricular tachycardia by automatic pulse generators sensing the atrial electrogram.
No takes yet. Share an insight, caveat, or question.
Barold et al. (1987) conducted a case report in Supraventricular tachycardia (n=2). Implanted atrial bipolar automatic burst tachycardia-terminating pulse generator (Intermedics CyberTach 60) was evaluated on Double sensing of P and QRS signals causing inappropriate delivery of burst stimulation. An implanted atrial bipolar automatic burst tachycardia-terminating pulse generator caused inappropriate burst stimulation due to double sensing of P and QRS signals during normal sinus rhythm.
Synapse has enriched 10 related papers. Consider them for comparative context: