Key result
Bipolar atrial sensing thresholds during induced atrial tachyarrhythmias were lower than during sinus rhythm, with a 63% reduction during atrial fibrillation.
Why the study?
How do bipolar atrial sensing thresholds during induced atrial tachyarrhythmias compare to those during sinus rhythm in patients with DDDR pacemakers?
Observational (n=43)
How do bipolar atrial sensing thresholds during induced atrial tachyarrhythmias compare to those during sinus rhythm in patients with DDDR pacemakers?
Atrial sensing thresholds are significantly reduced during atrial fibrillation compared to sinus rhythm, suggesting a 4:1 to 5:1 safety margin based on sinus rhythm is needed for adequate automatic mode switching.
Answers are grounded in this paper and cite their sources.
May support larger sinus-rhythm sensing margins for reliable AMS; leaves open prospective validation in spontaneous arrhythmias.
UNLABELLED: Automatic mode switching (AMS) function in dual chamber pacemakers depends on adequate detection of atrial tachyarrhythmias. There are few data on showing how intra-operative atrial signal amplititude during sinus rhythm can predict atrial tachyarrhythmias after pacemaker implantation. In 43 patients undergoing DDDR pacemaker implantation and atrioventricular nodal ablation for the treatment of drug-refractory paroxysmal atrial fibrillation, atrial sensing thresholds during sinus rhythm and during induced atrial tachyarrhythmias (24-48 h after device implantation) were analysed. Five different DDDR pacemaker systems were implanted (Chorus 7034, Ela Medical n = 13; Meta DDDR 1254, Telectronics Pacing Systems n = 12; Vigor DR 1230, Guidant n = 6; Trilogy DR 2364, Pacesetter, n = 2; Kappa DR 401, Medtronic USA n = 10). Every patient received a steroid-eluting, screwing, bipolar atrial lead (Medtronic, Capsure-Fix 4068). The mean P wave amplitude during implantation was 3.91 +/- 1.14 mV. The mean atrial sensing threshold during sinus rhythm and during all modes of induced atrial tachyarrhythmias was 3.35 +/- 1.0 mV, and 1.52 +/- 0.92 mV, respectively (P < 0.001). Atrial fibrillation was induced in 36 patients. The mean sensing threshold during sinus rhythm in this patient group was 3.39 +/- 1.01 mV, the mean sensing threshold during atrial fibrillation was 1.27 +/- 0.56 mV, reflecting a 63% reduction of sensing threshold compared with sinus rhythm (P < 0.001). Atrial flutter was induced in seven patients. The mean sensing threshold during sinus rhythm was 2.92 +/- 1.19 mV, the mean sensing threshold during atrial flutter was 2.79 +/- 1.26 mV, reflecting a reduction of 5% (ns) compared with sinus rhythm. Atrial sensing thresholds during sinus rhythm were significantly correlated with sensing thresholds during atrial tachyarrhythmias (r = 0.44; P < 0.002), but there were significant variations in intra-individual results. The reduction of atrial sensing thresholds between sinus rhythm and induced atrial tachyarrhythmias ranged from 30% to 82%. CONCLUSION: Bipolar atrial sensing thresholds during sinus rhythm are correlated with sensing thresholds during atrial tachyarrhythmias, but there is a large degree of variance in individual patients. A 4:1 to 5:1 atrial sensing safety margin based on sensing threshold during sinus rhythm is a predictor for adequate postoperative detection of atrial tachyarrhythmias and the function of AMS devices.
No takes yet. Share an insight, caveat, or question.
Jörg Neuzner (1999) conducted an observational in drug-refractory paroxysmal atrial fibrillation (n=43). Bipolar atrial sensing thresholds during induced atrial tachyarrhythmias vs. Bipolar atrial sensing thresholds during sinus rhythm was evaluated on Atrial sensing threshold (mV) (p=<0.001). Bipolar atrial sensing thresholds during induced atrial tachyarrhythmias were lower than during sinus rhythm, with a 63% reduction during atrial fibrillation.