Key result
Bifocal pacing boosts LV dp/dt max ~13% from baseline but remains inferior to biventricular pacing.
Why the study?
Does bifocal pacing improve left ventricular hemodynamics compared to baseline and biventricular pacing in patients with severe congestive heart failure?
Does bifocal pacing improve left ventricular hemodynamics compared to baseline and biventricular pacing in patients with severe congestive heart failure?
Bifocal pacing improves left ventricular hemodynamics compared to baseline, but biventricular pacing remains superior for cardiac resynchronization.
Answers are grounded in this paper and cite their sources.
Bifocal pacing may modestly improve hemodynamics in severe HF but is inferior to biventricular pacing; leaves open its clinical role pending randomized trials.
BACKGROUND: Bifocal pacing (BFP) has been proposed as a more feasible alternative of cardiac resynchronization therapy (CRT). AIM: To evaluate BFP in patients with severe congestive heart failure and significant intraventricular conduction delay and to compare it with biventricular pacing (BVP). METHODS: Both echocardiographic examination including tissue Doppler imaging and invasive measurements of cardiac hemodynamics was performed under basal conditions and during BFP and BVP. RESULTS: 50 patients were included: 29 patients with ischemic heart disease (IHD), 21 patients with idiopathic dilated cardiomyopathy (IDCM). Left ventricular (LV) pressure gradient (dp/dt max) increased during BFP compared to the baseline (13.4 %, 95 % CI 9.2-17.6 %, p < 0.0001) and a further increase was achieved during BVP (29.5 %, 95 % CI 23.7-35.4 %, p < 0.0001). A significant correlation was found between the distance of the right ventricular apical and outflow tract leads and percentage of dp/dt max increase in IDCM patients (r = 0.72, p < 0.001), but not in IHD patients. Interventricular mechanical delay (IVMD) decreased in BFP compared to baseline (43 +/- 22 ms vs. 53 +/- 31 ms, p = 0.006). BVP produced even shorter IVMD (22 +/- 19 ms, p < 0.0001). In all patients, the regional systolic contraction times were significantly shortened, corresponding with prolongation of the respective regional diastolic filling times during both BFP (p < 0.05 for all segments) and BVP (p < 0.001 for all segments). The effect of BVP on regional systole shortening was more pronounced. CONCLUSIONS: BFP improves LV hemodynamics by decreasing the inter- and intraventricular conduction delays. The leads in the right ventricle should be placed at the longest achievable distance. BVP is superior to BFP.
No takes yet. Share an insight, caveat, or question.
Bulava et al. (2006) studied Severe congestive heart failure with significant intraventricular conduction delay (n=50). Bifocal pacing vs. Baseline (intrinsic rhythm) and Biventricular pacing was evaluated on Percentage increase in left ventricular pressure gradient (dp/dt max) from baseline (13.4% increase, 95% CI 9.2-17.6, p=<0.0001). Bifocal pacing significantly increased left ventricular dp/dt max by 13.4% compared to baseline, although biventricular pacing provided a superior increase of 29.5%.
Synapse has enriched 10 related papers. Consider them for comparative context: