Key result
Higher initial T/QRS baseline height linked to more ST events during fetal monitoring.
Why the study?
Does the orientation of the fetal electrical heart axis affect the incidence of ST events during fetal monitoring in labour?
Observational (n=1,097)
Open-label
Yes
Does the orientation of the fetal electrical heart axis affect the incidence of ST events during fetal monitoring in labour?
Effect estimate: correlation coefficient 0.63
p-value: p=<0.001
The orientation of the fetal electrical heart axis significantly affects the incidence of ST events during fetal monitoring, which should be accounted for to reduce false alarms.
ST events increase with higher baseline T/QRS irrespective of fetal acidosis; challenges unadjusted ST monitoring and leaves open prospective validation of axis-adjusted thresholds.
OBJECTIVE: Reducing perinatal morbidity and mortality is one of the major challenges in modern health care. Analysing the ST segment of the fetal electrocardiogram was thought to be the breakthrough in fetal monitoring during labour. However, its implementation in clinical practice yields many false alarms and ST monitoring is highly dependent on cardiotocogram assessment, limiting its value for the prediction of fetal distress during labour. This study aims to evaluate the relation between physiological variations in the orientation of the fetal electrical heart axis and the occurrence of ST events. METHODS: A post-hoc analysis was performed following a multicentre randomised controlled trial, including 1097 patients from two participating centres. All women were monitored with ST analysis during labour. Cases of fetal metabolic acidosis, poor signal quality, missing blood gas analysis, and congenital heart disease were excluded. The orientation of the fetal electrical heart axis affects the height of the initial T/QRS baseline, and therefore the incidence of ST events. We grouped tracings with the same initial baseline T/QRS value. We depicted the number of ST events as a function of the initial baseline T/QRS value with a linear regression model. RESULTS: A significant increment of ST events was observed with increasing height of the initial T/QRS baseline, irrespective of the fetal condition; correlation coefficient 0.63, p<0.001. The most frequent T/QRS baseline is 0.12. CONCLUSION: The orientation of the fetal electrical heart axis and accordingly the height of the initial T/QRS baseline should be taken into account in fetal monitoring with ST analysis.
No takes yet. Share an insight, caveat, or question.
Vullings et al. (2017) conducted an observational in Fetal monitoring during labour (n=1,097). Initial baseline T/QRS value vs. Lower baseline T/QRS values was evaluated on Incidence of ST events (correlation coefficient 0.63, p=<0.001). Increasing height of the initial T/QRS baseline was significantly associated with a higher incidence of ST events during fetal monitoring, irrespective of fetal condition (correlation coefficient 0.63, p<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: