Key result
Percutaneous mitral valved stent implantation achieves 100% in vitro success with comparable fluoroscopy and endoscopy times.
Why the study?
Is a novel percutaneous transapical system for transcatheter mitral valve implantation feasible in an in vitro porcine model?
Is a novel percutaneous transapical system for transcatheter mitral valve implantation feasible in an in vitro porcine model?
A novel percutaneous transapical system for transcatheter mitral valve implantation demonstrated feasibility and excellent apical fixation in an in vitro porcine model.
Hybrid access for mitral valved stent implantation expands fixation options; leaves open clinical safety and efficacy.
Objectives: Despite recent achievements, implantation of a transcatheter mitral valved stent remains challenging. In this study, we present a different approach for implantation of a percutaneous mitral valved stent. Methods: Percutaneous transapical access is combined with, respectively, a left-transatrial, right-transatrial/transseptal or transfemoral/transseptal approach for mitral valve stent implantation and secure fixation. The apical fixation and occlusion are ensured with an Amplatzer occluder. This novel approach was tested in 22 porcine hearts in an in vitro setting under the guidance of fluoroscopy ( n = 11) and endoscopy ( n = 11). The in vitro setup included continuous flushing at 37 °C. We determined the feasibility, time of implantation, stent deployment and stent fixation. Results: Percutaneous mitral valved stent implantation was successful in all cases. Good handling properties and precise positioning were achieved. Time of implantation was comparable in the fluoroscopic and endoscopic groups at 10:41 ± 3:18 and 10:09 ± 2:42 min, respectively. Apical fixation with the occluder was excellent in all 22 cases. Conclusions: The feasibility of percutaneous mitral valved stent implantation has been demonstrated in preliminary in vitro experiments. Subsequent studies are warranted to determine the efficacy of this minimally invasive catheter-based mitral valved stent implantation.
No takes yet. Share an insight, caveat, or question.
Dai et al. (2017) studied Mitral valve stent implantation (n=22). Percutaneous transapical mitral valved stent implantation vs. Fluoroscopy vs Endoscopy guidance was evaluated on Feasibility, time of implantation, stent deployment and stent fixation. Percutaneous mitral valved stent implantation was successful in all 22 in vitro porcine hearts, with comparable implantation times under fluoroscopy (10:41 min) and endoscopy (10:09 min).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: