Key result
Single-lead ECG screening using Eko DUO in neonates yielded interpretable recordings in 63% of cases, but QT interval measurement was mostly not possible.
Why the study?
Neonatal ECG screening can identify congenital long QT syndrome and other heart diseases, but obtaining a 12-lead ECG in infants is difficult.
Is a single-lead ECG recorded with a handheld device (Eko DUO) feasible and interpretable for screening in neonates?
Cross-Sectional (n=100)
Is a single-lead ECG recorded with a handheld device (Eko DUO) feasible and interpretable for screening in neonates?
Single-lead ECG screening using a handheld device in neonates is feasible for assessing rate and rhythm but lacks the fidelity required for reliable QT interval measurement.
Answers are grounded in this paper and cite their sources.
Single-lead ECG screening feasibility remains limited in neonates; hypothesis-generating and should not yet change practice.
Background: Neonatal electrocardiogram (ECG) screening can potentially identify congenital long QT syndrome and other heart diseases. Early identification is likely to reduce mortality. A barrier is the difficulty in obtaining a 12-lead ECG in an infant. We aimed to assess the feasibility of using a single lead ECG recorded with a handheld device (Eko DUO) in neonates. Methods: This cross-sectional study included neonates in the first 3 days of life. We recorded a single-lead ECG using the Eko DUO. Time from the beginning of recording until a satisfactory recording was measured with a stopwatch and reported as mean and standard deviation. We reported the ratio of interpretable recordings and the proportion of those with any abnormalities. Results: The mean time for recording was 198.1 ± 94.7 s. The total number of interpretable recordings was 63% ( n = 63). Of the interpretable recordings, one neonate was found to have tachycardia (1.6%). Conclusions: Neonatal ECG screening using Eko DUO is feasible. However, the fidelity of the recorded ECG is suboptimal. While most of the ECGs were interpretable for rate and rhythm, QT interval measurement was not possible in most neonates.
No takes yet. Share an insight, caveat, or question.
Selvaraj et al. (2024) conducted a cross-sectional in Neonatal heart disease screening (n=100). Single-lead ECG using Eko DUO was evaluated on Ratio of interpretable recordings. Single-lead ECG screening using Eko DUO in neonates yielded interpretable recordings in 63% of cases, but QT interval measurement was mostly not possible.