Programming changes resolve unusual double cross-talk in a dual-chamber ICD case.
Why the study?
Far-field ventricular events are commonly sensed in the atrial channel, but the reverse is uncommon and observing cross-talk in both chambers simultaneously had never been reported.
Case Report (n=1)
This case report identifies a previously unreported phenomenon of double cross-talk in a dual-chamber ICD and provides a programming solution to resolve it.
Answers are grounded in this paper and cite their sources.
May prompt recognition of rare double cross-talk in dual-chamber devices; leaves open optimal programming strategies pending larger studies.
Cross-talk is a well-described phenomenon in a dual-chamber cardiovascular implantable electronic device. Far-field ventricular events are more commonly sensed in the atrial channel, the reverse is uncommon, and seeing both at the same time has never been reported. We present a case of double cross-talk in a dual-chamber Medtronic ® implantable cardioverter-defibrillator. In this report, we decipher an unusual device response to the cross-talk and describe the programming changes required to resolve it.
No takes yet. Share an insight, caveat, or question.
Almehmadi et al. (2020) conducted a case report in Double cross-talk in a dual-chamber implantable cardioverter-defibrillator (n=1). Programming changes was evaluated on Resolution of double cross-talk. Programming changes successfully resolved an unusual case of double cross-talk in a dual-chamber Medtronic implantable cardioverter-defibrillator.