Key result
Programming changes resolve unusual double cross-talk in a dual-chamber ICD case.
Why the study?
Far-field ventricular events are commonly sensed in the atrial channel, but the reverse is uncommon and observing cross-talk in both chambers simultaneously had never been reported.
Population
A case of a patient with a dual-chamber Medtronic implantable cardioverter-defibrillator
Design
Case report
Authors
Loading...
May prompt recognition of rare double cross-talk in dual-chamber devices; leaves open optimal programming strategies pending larger studies.
Case Report (n=1)
This case report identifies a previously unreported phenomenon of double cross-talk in a dual-chamber ICD and provides a programming solution to resolve it.
Almehmadi et al. (2020) conducted a case report in Double cross-talk in a dual-chamber implantable cardioverter-defibrillator (n=1). Programming changes was evaluated on Resolution of double cross-talk. Programming changes successfully resolved an unusual case of double cross-talk in a dual-chamber Medtronic implantable cardioverter-defibrillator.