Key result
Novel oval and D-shaped transcatheter mitral valve stents demonstrate sufficient fixation and durability in vitro.
Why the study?
Do novel oval-shaped and D-shaped valved stents demonstrate sufficient fixation force and valve performance compared to an established reference valve in an in vitro model?
Do novel oval-shaped and D-shaped valved stents demonstrate sufficient fixation force and valve performance compared to an established reference valve in an in vitro model?
Novel oval-shaped and D-shaped valved stents for transcatheter mitral valve implantation demonstrate sufficient annular fixation force and adequate durability in an in vitro model.
Preclinical fixation data support further development of noncircular mitral stents; leaves open in vivo validation before any clinical consideration.
OBJECTIVES: In this study, results of a functional in vitro study of 2 newly developed valved stents for transcatheter mitral valve implantation are presented. METHODS: Two novel stent designs, an oval-shaped and a D-shaped stent with a strut fixation system were developed. The fixation force of the novel stents were tested in vitro in porcine hearts with a tensile test set-up. In further experiments, the stents were equipped with a circular valved stent, and the valve performances were investigated in a pulsatile heart valve tester. RESULTS: Sufficient mean stent fixation forces in the range of 24.2 ± 0.9 N to 28.6 ± 1.9 N were measured for the different stent models. The novel valved stents showed good performance in an in vitro pulsatile heart valve tester. A sufficient opening area and low opening pressures were measured for all tested mitral valved stents. Compared with an established reference valve, the D-shaped stent and the oval-shaped valved stent showed a lower systolic transvalvular pressure gradient, which indicates slightly greater extent of valvular leakage of the closed valved stents. However, the mitral nitinol valved stents demonstrated adequate durability. CONCLUSIONS: This study indicates a sufficient annular fixation force of the tested transcatheter mitral valve implantation valved stent prototypes. Therefore, these mitral valved stents demonstrate a new type of mitral valved stent design.
No takes yet. Share an insight, caveat, or question.
Loger et al. (2017) studied Transcatheter mitral valve implantation. Oval-shaped and D-shaped valved stents vs. Established reference valve was evaluated on Fixation force and valve performance. Novel oval-shaped and D-shaped valved stents for transcatheter mitral valve implantation demonstrated sufficient annular fixation forces (24.2-28.6 N) and adequate durability in an in vitro model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: