Key result
Abrupt ICD lead failure causing sudden death highlights the critical importance of device algorithm programming.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Highlights the critical importance of appropriate ICD algorithm programming to prevent sudden death in the event of catastrophic lead failure.
Implantable cardioverter defibrillator (ICD) therapy is commonly used for the prevention of sudden arrhythmic death in high risk populations. A recognized complication of such therapy is lead failure, leading to potential inappropriate detection and delivery of high energy defibrillation shocks. Many modern devices utilize algorithms which can differentiate between lead noise and true arrhythmia, avoiding unnecessary therapies. However, the programming of these algorithms can be the difference between life and death.
No takes yet. Share an insight, caveat, or question.
Tuohy et al. (2015) conducted a case report in Implantable cardioverter defibrillator lead failure (n=1). Implantable cardioverter defibrillator (ICD) therapy was evaluated. A single case of abrupt catastrophic implantable cardioverter defibrillator lead failure resulted in sudden death, highlighting the critical importance of device algorithm programming.