Key result
Portico TAVI via the FlexNav system yields 0% early major complications in a small series.
Why the study?
Despite limited initial data, uncertainty persists regarding the actual role of the novel FlexNav delivery system for Portico valve TAVI in real-world practice.
Is the novel FlexNav delivery system safe and feasible for transcatheter aortic valve implantation with the Portico device in patients with severe symptomatic aortic stenosis?
Observational (n=18)
No
Is the novel FlexNav delivery system safe and feasible for transcatheter aortic valve implantation with the Portico device in patients with severe symptomatic aortic stenosis?
The novel FlexNav delivery system for TAVI with the Portico valve appears feasible and safe, with no deaths, strokes, or major vascular complications in a small initial case series.
Small series supports FlexNav feasibility for Portico TAVI; hypothesis-generating and should not yet change practice.
BACKGROUND: Transcatheter aortic valve implantation (TAVI) is a safe and beneficial treatment for patients with severe symptomatic aortic stenosis at high and intermediate surgical risk. Success rates for TAVI continue to increase due to more refined procedural approaches and devices, and above all, to the improvements in procedural knowledge by TAVI operators. The development of a new delivery system represents an important evolutionary process in the TAVI procedure, both for the type of device and for the operators, as it may substantially expand indications, success rates, and safety. FlexNav (Abbott Cardiovascular) is a novel delivery system designed to improve ease and precision of TAVI using the Abbott Portico valve. Despite limited results on its advantages, such as those provided in the PORTICO investigational device exemption substudy, uncertainty persists on the actual role of FlexNav in real-world practice. METHODS AND RESULTS: We hereby present our preliminary experience with FlexNav in a consecutive series of 18 patients undergoing TAVI with Portico valve in early 2020 (3 with axillary access). Procedural results were excellent, as shown by the absence of deaths, strokes, and major vascular complications (95% confidence interval, 0%-18.5%), and pacemaker implantation in 3 subjects (21.4%; 95% confidence interval, 4.7%-50.8%). CONCLUSIONS: Despite the small sample size, the present case series suggests that adoption of the FlexNav novel delivery system for TAVI with Portico valve is feasible and safe, and holds the promise of further improving early and long-term results of this procedure.
No takes yet. Share an insight, caveat, or question.
Corcione et al. (2021) conducted an observational in Severe symptomatic aortic stenosis (n=18). FlexNav delivery system for TAVI with Portico valve was evaluated on 30-day death, stroke, major vascular complications, and major bleeding (95% CI 0-18.5). In a case series of 18 patients, TAVI with the Portico valve using the novel FlexNav delivery system resulted in no deaths, strokes, major vascular complications, or major bleeding at 30 days.