Key result
Apixaban use in an HVAD patient is linked to pump thrombosis requiring pump exchange.
Why the study?
Warfarin remains the agent of choice for thromboembolic prevention in LVAD patients, and the outcome of using apixaban in this population required exploration.
Does apixaban increase the risk of pump thrombosis in patients with left ventricular assist devices?
Case Report (n=1)
Does apixaban increase the risk of pump thrombosis in patients with left ventricular assist devices?
The use of apixaban in patients with left ventricular assist devices may lead to pump thrombosis, reinforcing the preference for warfarin in this population.
May signal apixaban-related pump thrombosis risk in HVAD; hypothesis-generating and should not yet change practice.
BACKGROUND Direct oral anticoagulants (DOAC) are currently the preferred agents for long-term anticoagulation in the appropriate patient with venous thromboembolism, non-valvular atrial fibrillation, and left ventricular thrombi because of their ease of use, fixed dosing, lack of need for routine monitoring, and limited dietary and drug interactions. However, warfarin is still the agent of choice for preventing thromboembolic events in patients with left ventricular assist devices (LVAD). In this case report, we explore the outcome of using apixaban in a patient with an LVAD. CASE REPORT A 56-year-old woman with morbid obesity and stage D congestive heart failure status after HeartWare ventricular assist device (HVAD) placement 2 years prior, who was on long-term anticoagulation with apixaban after failure of warfarin therapy, presented to the Emergency Department with 2 months of worsening fatigue, dark urine, and 1 day of low-flow alarms from her HVAD. Laboratory and radiographic data were consistent with a diagnosis of pump thrombosis. She underwent pump exchange and was started on a heparin drip. Genetic testing for warfarin resistance was negative. Detailed history-taking revealed that the failure to maintain a therapeutic international normalized ratio (INR) was likely due to dietary factors. She was re-challenged with warfarin, and a therapeutic INR level was reached shortly after initiation. She was later discharged on a stable dose of warfarin and remained in a good clinical state without any major adverse events at the 1-year follow-up. CONCLUSIONS Apixaban can be associated with an increased risk of thrombosis in patients with HVADs and should be used with caution and only in select patients.
No takes yet. Share an insight, caveat, or question.
A 2021 study conducted a case report in Congestive heart failure with left ventricular assist device (LVAD) (n=1). Apixaban was evaluated on Pump thrombosis. Apixaban use in a 56-year-old woman with a HeartWare ventricular assist device was associated with the development of pump thrombosis requiring pump exchange.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: