Key result
Posterior cardiac vein auxiliary shock before primary shock reduces mean DFT energy by ~61%.
Why the study?
Does an auxiliary shock delivered to the left posterior coronary vein before a primary endocardial shock reduce the defibrillation threshold in dogs?
Does an auxiliary shock delivered to the left posterior coronary vein before a primary endocardial shock reduce the defibrillation threshold in dogs?
Addition of an auxiliary shock from the posterior cardiac vein before an endocardial primary shock reduces defibrillation threshold energy by more than 50% in a canine model.
Answers are grounded in this paper and cite their sources.
May reduce defibrillation thresholds in dogs; hypothesis-generating for human ICD configurations, requiring prospective trials.
INTRODUCTION: For endocardial shocks near the defibrillation threshold (DFT), postshock activity originates from the lateral left ventricular apex, where the shock field is weak. This study tested the hypothesis that an auxiliary shock (AS) delivered between an electrode at this site and a superior vena cava (SVC) electrode before the primary endocardial shock (PS) would reduce the DFT. METHODS AND RESULTS: In six pentobarbital-anesthetized dogs (26 to 36 kg), catheter electrodes were placed in the right ventricular (RV) apex and the SVC. To simulate transvenous introduction, a small electrode was inserted into the posterior cardiac vein using an epicardial approach. For dual shock treatments, AS (2-msec monophasic) was applied to the coronary vein electrode at different time intervals before a biphasic PS (4 msec/3 msec) to the RV-SVC electrodes. The mean DFT energy for dual shocks treatments were significantly reduced (P < 0.05) in comparison to the control treatment (no AS, 26.5+/-8.8 J). Mean DFT energy after 10 seconds of electrically induced ventricular fibrillation for dual shocks, in which AS and PS were separated by 1, 5, 10, and 20 msec, were 10.2+/-4.1 J, 10.9+/-5.5 J, 11.3+/-6.3 J, and 15.4+/-7.2 J, respectively. These values were all significantly lower than the PS alone (26.5+/-8.8 J). CONCLUSION: Addition of an AS from the posterior cardiac vein before an endocardial PS reduces DFT energy by more than 50%. Such DFT reduction could improve therapeutic safety margin or permit reduction in volume of implantable cardioverter defibrillators.
No takes yet. Share an insight, caveat, or question.
KenKnight et al. (2000) studied Ventricular fibrillation (n=6). Auxiliary shock before primary endocardial shock vs. Primary endocardial shock alone was evaluated on Mean defibrillation threshold (DFT) energy (p=<0.05). Addition of an auxiliary shock from the posterior cardiac vein before an endocardial primary shock reduced mean defibrillation threshold energy from 26.5 J to as low as 10.2 J.
Synapse has enriched 10 related papers. Consider them for comparative context: