We studied the effects of small, < 20 mum, Te inclusions on the energy resolution of CdZnTe gamma-ray detectors using a highly collimated X-ray beam and gamma-rays, and modeled them via a simplified geometrical approach. Previous reports demonstrated that Te inclusions of about a few microns in diameter degraded the charge-transport properties and uniformity of CdZnTe detectors. The goal of this work was to understand the extent to which randomly distributed Te-rich inclusions affect the energy resolution of CZT detectors, and to define new steps to overcome their deleterious effects. We used a phenomenological model, which depends on several adjustable parameters, to reproduce the experimentally measured effects of inclusions on energy resolution. We also were able to bound the materials-related problem and predict the enhancement in performance expected by reducing the size and number of Te inclusions within the crystals.
No takes yet. Share an insight, caveat, or question.
Bolotnikov et al. (2007) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: