Key result
CorVue algorithm fails to reliably predict heart failure events.
Why the study?
The long-term efficacy of the CorVue algorithm to predict heart failure hospitalizations in a real-life cohort had not been systematically evaluated.
Does the CorVue intrathoracic impedance algorithm accurately predict heart failure events in patients with ICD/CRT-D devices?
Observational (n=53)
Does the CorVue intrathoracic impedance algorithm accurately predict heart failure events in patients with ICD/CRT-D devices?
The CorVue intrathoracic impedance algorithm showed low sensitivity and positive predictive value for predicting heart failure events in a real-world cohort, suggesting routine activation may generate misleading information.
Low CorVue sensitivity and high false negatives caution against routine activation; leaves open whether algorithm refinements or multimodality monitoring enhance HF prediction.
BACKGROUND: Heart failure (HF) hospitalizations have a negative impact on quality of life and imply important costs. Intrathoracic impedance (ITI) variations detected by cardiac devices have been hypothesized to predict HF hospitalizations. Although Optivol™ algorithm (Medtronic, Minneapolis, MN, USA) has been widely studied, CorVue™ algorithm's (St. Jude Medical, St. Paul, MN, USA) long-term efficacy has not been systematically evaluated in a "real-life" cohort. METHODS: CorVue™ was activated in implantable cardioverter defibrillator (ICD)/cardiac resynchronization therapy defibrillator (CRT-D) patients to store information about ITI measures. Clinical events (new episodes of HF requiring treatment and hospitalizations) and CorVue™ data were recorded every 3 months. Appropriate CorVue™ detection for HF was considered if it occurred in the 4 prior weeks to the clinical event. RESULTS: Fifty-three ICD/CRT-D (26 ICD and 27 CRT-D) patients (67 ± 1 years old, 79% male) were included. Device position was subcutaneous in 28 patients. At inclusion, mean left ventricular ejection fraction was 25 ± 7% and 27 patients (51%) were in New York Heart Association class I, 18 (34%) in class II, and eight (15%) in class III. After a mean follow-up of 17 ± 9 months, 105 ITI drops alarms were detected in 32 patients (60%). Only six alarms were appropriate (true positive) and required hospitalization. Eighteen patients (34%) presented 25 clinical episodes (12 hospitalizations and 13 emergency room/ambulatory treatment modifications). Nineteen of these clinical episodes (76%) remained undetected by the CorVue™ (false negative). Sensitivity of CorVue™ resulted in 24%, specificity was 70%, positive predictive value of 6%, and negative predictive value of 93%. CONCLUSIONS: CorVue™ showed a low sensitivity to predict HF events. Therefore, routinely activation of this algorithm could generate misleading information.
No takes yet. Share an insight, caveat, or question.
Palfy et al. (2018) conducted an observational in Heart failure (n=53). CorVue algorithm was evaluated on Prediction of heart failure events (new episodes requiring treatment and hospitalizations). The CorVue algorithm demonstrated low efficacy for predicting heart failure events, yielding a sensitivity of 24%, specificity of 70%, and positive predictive value of 6%.