New possibilities are opened up utilizing position-sensitive germanium detectors in the X-ray spectroscopy of highly charged heavy ions at GSI-Darmstadt. The recent experiments revealed the need for two-dimensional strip detectors with their inherent advantages concerning spectroscopy and imaging capabilities as well as polarization sensitivity. For the first prototype a germanium diode (70 mm/spl times/41 mm, 11 mm thick) with a boron implanted contact and an amorphous Ge contact was prepared. A 128-strip structure on an area of 32 mm/spl times/56 mm with a pitch of 250 /spl mu/m on the front contact (implanted) and a 48-strip structure with a pitch of 1167 /spl mu/m on the rear contact (amorphous Ge) are realized with the help of plasma etching. The detector is mounted in a cryostat which will enable any orientation of the detector with respect to a photon source. The results of laboratory tests and planned applications at the ESR storage ring in Darmstadt are presented.
No takes yet. Share an insight, caveat, or question.
Protić et al. (2005) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: