Key result
Deep learning algorithm matches radiologists in ~87% of thoracic aorta measurements and halves analysis time.
Why the study?
Manual diameter measurements on ECG-gated CTA for thoracic aortic dilatation are the gold standard but are time-consuming and show high inter-reader variability.
Does a deep learning algorithm maintain accuracy and reduce measurement time compared to manual radiologist measurements for thoracic aorta diameters on ECG-gated CTA in patients with suspected aortic dilatation?
Observational (n=371)
Does a deep learning algorithm maintain accuracy and reduce measurement time compared to manual radiologist measurements for thoracic aorta diameters on ECG-gated CTA in patients with suspected aortic dilatation?
A deep learning algorithm can perform automated, guideline-compliant thoracic aorta diameter measurements on CTA with high coherence to radiologists while saving significant time.
Answers are grounded in this paper and cite their sources.
May accelerate aortic measurements in CTA workflows; leaves open prospective validation before clinical adoption.
BACKGROUND: Manually performed diameter measurements on ECG-gated CT-angiography (CTA) represent the gold standard for diagnosis of thoracic aortic dilatation. However, they are time-consuming and show high inter-reader variability. Therefore, we aimed to evaluate the accuracy of measurements of a deep learning-(DL)-algorithm in comparison to those of radiologists and evaluated measurement times (MT). METHODS: We retrospectively analyzed 405 ECG-gated CTA exams of 371 consecutive patients with suspected aortic dilatation between May 2010 and June 2019. The DL-algorithm prototype detected aortic landmarks (deep reinforcement learning) and segmented the lumen of the thoracic aorta (multi-layer convolutional neural network). It performed measurements according to AHA-guidelines and created visual outputs. Manual measurements were performed by radiologists using centerline technique. Human performance variability (HPV), MT and DL-performance were analyzed in a research setting using a linear mixed model based on 21 randomly selected, repeatedly measured cases. DL-algorithm results were then evaluated in a clinical setting using matched differences. If the differences were within 5 mm for all locations, the cases was regarded as coherent; if there was a discrepancy >5 mm at least at one location (incl. missing values), the case was completely reviewed. RESULTS: HPV ranged up to ±3.4 mm in repeated measurements under research conditions. In the clinical setting, 2,778/3,192 (87.0%) of DL-algorithm's measurements were coherent. Mean differences of paired measurements between DL-algorithm and radiologists at aortic sinus and ascending aorta were -0.45±5.52 and -0.02±3.36 mm. Detailed analysis revealed that measurements at the aortic root were over-/underestimated due to a tilted measurement plane. In total, calculated time saved by DL-algorithm was 3:10 minutes/case. CONCLUSIONS: The DL-algorithm provided coherent results to radiologists at almost 90% of measurement locations, while the majority of discrepent cases were located at the aortic root. In summary, the DL-algorithm assisted radiologists in performing AHA-compliant measurements by saving 50% of time per case.
No takes yet. Share an insight, caveat, or question.
Pradella et al. (2021) conducted an observational in suspected aortic dilatation (n=371). Deep learning algorithm vs. Manual measurements by radiologists was evaluated on Coherence of measurements (difference within 5 mm) between DL-algorithm and radiologists. A deep learning algorithm provided coherent thoracic aorta diameter measurements compared to radiologists in 87.0% of locations and saved 50% of measurement time per case.