Key result
Real-time LVAD cardiac output estimation achieves >0.99 correlation with true stroke volume in experimental models.
Why the study?
A low-invasive, real-time method for estimating cardiac output in circulatory systems supported by LVADs is needed for clinical application.
Does a new real-time estimation method accurately estimate cardiac output in a circulatory system supported by an LVAD?
Does a new real-time estimation method accurately estimate cardiac output in a circulatory system supported by an LVAD?
Effect estimate: r >0.99 or 0.95
A novel, minimally invasive method for real-time estimation of cardiac output during LVAD support showed high correlation with true stroke volume in mock and animal models.
May facilitate real-time cardiac output monitoring in LVAD support; leaves open clinical validation before practice change.
A new method for indirect and real-time estimation of the cardiac output of the circulatory system supported by the left ventricular assist device (LVAD) was proposed. This method has low invasiveness and is useful for clinical application of the LVAD since it needs only two measurements: the outflow rate of blood from the LVAD and the aortic pressure. The proposed method is based on a system identification technique for the time series model of the cardiovascular system and requires less computational time than other methods with similar estimation accuracy. Hence, the proposed method could be implemented in a personal computer system and realize the on-line and real-time estimation of the instantaneous outflow rate of the natural heart. The efficiency and limitation of the method were discussed on the basis of the results obtained from in vitro experiment using a mock circulatory system and in vivo experiment using an adult goat. It was revealed that the method could obtain a fairly high correlation coefficient between the true stroke volume of the natural heart and its estimate of more than 0.99 (in vivo) or 0.95 (in vivo). This means that there is a fair possibility of applying the proposed estimation system to the clinical use.
No takes yet. Share an insight, caveat, or question.
Yoshizawa et al. (1993) studied Circulatory system supported by a left ventricular assist device (LVAD). Real-time cardiac output estimation method was evaluated on Correlation coefficient between the true stroke volume of the natural heart and its estimate (r >0.99 or 0.95). A new real-time cardiac output estimation method for LVAD-supported systems achieved a correlation coefficient of >0.99 or 0.95 with the true stroke volume in experimental models.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: