Key result
Gore Helex and Cardioform PFO devices linked to pericardial tamponade from wire frame fracture.
Case Report (n=2)
This report provides the first published evidence of atrial erosion and pericardial tamponade associated with Gore PFO closure devices due to wire fracture.
Answers are grounded in this paper and cite their sources.
No published atrial erosion with Gore PFO occluder may inform device selection; leaves open long-term comparative safety data.
Percutaneous patent foramen ovale (PFO) closure is recommended for secondary prevention of paradoxical embolism through a PFO. In the United States, two Food and Drug Administration-approved PFO closure devices are currently available, and the choice depends on operator preference and PFO anatomy. Although these devices are easy to implant, there are several potential complications. As opposed to the Amplatzer PFO Occluder, there has been no published case of atrial erosion with Gore closure devices. This report describes two cases of pericardial tamponade due to perforation of the atrial wall induced by a wire frame fracture of the Gore Helex and Cardioform devices.
No takes yet. Share an insight, caveat, or question.
Kumar et al. (2019) conducted a case report in Patent foramen ovale (n=2). Gore Helex and Cardioform devices was evaluated on Pericardial tamponade due to atrial wall perforation. Gore Helex and Cardioform PFO closure devices can cause pericardial tamponade due to perforation of the atrial wall induced by a wire frame fracture.
Synapse has enriched 10 related papers. Consider them for comparative context: