Key result
Fractured HELEX Septal Occluder frame perforates anterior mitral valve leaflet causing significant regurgitation requiring surgery.
Case Report (n=1)
This case report highlights a severe complication of HELEX Septal Occluder frame fracture leading to mitral valve perforation, an event not previously associated with deleterious clinical effects.
Answers are grounded in this paper and cite their sources.
Highlights mitral perforation risk from HELEX frame fracture; leaves open need for enhanced device surveillance.
One or more breaks in the wire frame of the HELEX Septal Occluder (W.L. Gore and Associates, Flaggstaff, Arizona) were identified in 5% of patients in the multicenter pivotal trial leading to FDA approval. Fractures of the perimeter frame have not previously been associated with any apparent deleterious clinical effects. We report a case in which the broken ends of wire at a frame fracture were located in a position that resulted in perforation of the anterior mitral valve leaflet. There was significant mitral regurgitation and the device was removed surgically. The possible etiologies and prevention of wire frame fractures are also discussed.
No takes yet. Share an insight, caveat, or question.
Qureshi et al. (2009) conducted a case report in HELEX Septal Occluder frame fracture (n=1). HELEX Septal Occluder was evaluated on Mitral valve perforation and regurgitation. A fractured wire frame of a HELEX Septal Occluder resulted in perforation of the anterior mitral valve leaflet and significant mitral regurgitation, necessitating surgical removal.