Key result
Navitor TAVI in small annuli shows similar 12-month major event rates versus non-small annuli.
Why the study?
Does aortic annulus size (small vs non-small) impact clinical and hemodynamic outcomes in patients undergoing TAVI with the Navitor valve?
Cohort (n=218)
Does aortic annulus size (small vs non-small) impact clinical and hemodynamic outcomes in patients undergoing TAVI with the Navitor valve?
TAVI with the Navitor valve provides comparable mid-term clinical outcomes and bioprosthetic valve function in patients with small versus non-small aortic annuli, despite slightly higher but clinically acceptable post-procedural gradients in the small annulus group.
Answers are grounded in this paper and cite their sources.
Supports Navitor TAVI feasibility in small annuli; leaves open randomized confirmation of equivalence.
Background Patients with small aortic annuli undergoing transcatheter aortic valve implantation (TAVI) are at risk of higher residual transvalvular gradients. Evidence on the performance of new‐generation intra‐annular self‐expanding valves in this anatomical setting remains limited. Aims To evaluate the impact of annular size on procedural, hemodynamic, and mid‐term clinical outcomes after TAVI with the Navitor valve. Methods Patients were stratified by annular area measured by multi‐slice computed tomography, defining a small annulus as ≤ 430 mm 2 . Primary endpoints at 12 months were: (i) a clinical composite of all‐cause death, disabling stroke, or heart failure rehospitalization, and (ii) a composite endpoint of bioprosthetic valve dysfunction. Secondary endpoints included VARC‐3–defined procedural and early outcomes and valve hemodynamic performance. Results The study population comprised 218 patients, of whom 91 (41.7%) had a small annulus. VARC‐3 technical success was high and comparable between small and non‐small annuli (96.7% vs. 96.1%). Patients with small annuli had higher early post‐procedural mean transvalvular gradients (7.7 ± 3.8 vs. 6.6 ± 3.0 mmHg; p = 0.029) and smaller effective orifice areas index (1.1 ± 0.3 vs. 1.3 ± 0.2 cm 2 /m 2 ; p = 0.037), while the incidence of mean gradients ≥ 20 mmHg remained low (1.1% vs. 0.0%; p = 0.432). At 12 months, the clinical composite endpoint occurred in 8.8% of patients with small annuli and 10.2% with non‐small annuli ( p = 0.759). Bioprosthetic valve dysfunction was infrequent and similar between groups (6.6% vs. 6.3%; p = 0.875). Conclusions TAVI with the Navitor intra‐annular self‐expanding valve was associated with high procedural success, favorable valve hemodynamics, and comparable mid‐term clinical outcomes irrespective of annular size.
No takes yet. Share an insight, caveat, or question.
Casenghi et al. (2026) conducted a cohort in Aortic stenosis requiring transcatheter aortic valve implantation (n=218). Small aortic annulus (≤ 430 mm2) vs. Non-small aortic annulus (> 430 mm2) was evaluated on Clinical composite of all-cause death, disabling stroke, or heart failure rehospitalization at 12 months (p=0.759). TAVI with the Navitor valve in small aortic annuli resulted in similar 12-month rates of death, stroke, or heart failure rehospitalization compared to non-small annuli.
Synapse has enriched 10 related papers. Consider them for comparative context: