Key result
Pacemaker and ICD infections are linked to ~11% 30-day mortality, driven predominantly by staphylococci.
Why the study?
Infections associated with permanent pacemakers and ICD devices have a broader clinical spectrum than often reported, requiring elucidation of clinical features, causative microorganisms, management, and outcomes.
Observational (n=91)
Infections associated with permanent pacemakers and ICDs have a high short-term mortality (11% at 30 days) but a low risk of recurrence following device extraction and antibiotic therapy.
Device infections carry high short-term mortality; supports targeted prevention but leaves open optimal management in prospective trials.
Endocarditis and localized pocket infections are recognized as serious adverse events in patients with implanted cardiac impulse generators. We have undertaken a 10-y retrospective study in North Denmark Region (population 0.5 million) in order to elucidate the clinical spectrum, causative microorganisms, management and outcome. Infections associated with permanent pacemakers (PPM) and implanted cardioverter-defibrillator (ICD) devices were identified by searching hospital databases. Ninety-one incident cases were recorded in 1999 through 2008: 26 patients had endocarditis, 39 patients had a localized pocket infection, and 9 patients developed surgical sepsis with or without local signs immediately after implantation or reoperation; the device was the likely but unconfirmed focus of infection in 17 patients with bacteraemia. Staphylococcus aureus, coagulase-negative staphylococci and other Gram-positive bacteria were the predominant causative agents; only 6 cases were culture-negative. Management included device and lead extraction and individualized antibiotic therapy. The all-cause 30-day case-fatality was 11%. Only 3 recurrences were recorded during 2 y of follow-up. In conclusion, infections associated with permanent impulse generators have a broader clinical spectrum than often reported in the literature. Most cases are culture-positive with staphylococcal predominance. The short-term mortality is notably high, but the risk of recurrence is low.
No takes yet. Share an insight, caveat, or question.
Smit et al. (2010) conducted an observational in Infections associated with permanent pacemakers and implanted cardioverter-defibrillator devices (n=91). Permanent pacemakers (PPM) and implanted cardioverter-defibrillator (ICD) devices was evaluated on All-cause 30-day case-fatality. Infections associated with permanent pacemakers and ICDs were predominantly caused by staphylococci and resulted in an all-cause 30-day case-fatality of 11%, with only 3 recurrences over 2 years.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: