A hybrid composite based on Urushibara nickel-A and reduced graphene oxide (U-Ni-A/rGO) was developed for nonenzymatic glucose sensing in alkaline media. The composite combines skeletal nickel with a conductive rGO framework, giving an rGO-supported Ni interface with improved dispersion and lower interfacial resistance than U-Ni-A under identical conditions. The electrochemical response was examined by cyclic voltammetry (CV) and chronoamperometry (CA). CV was used to probe concentration-dependent behavior across 0.01–6000 μM, yielding a concentration-specific CV-derived slope of 5,666,475 μA mM −1 cm −2 in the 0.01–0.1 μM window. Quantitative sensing performance was evaluated by CA at 480 mV vs SCE, giving a linear response from 1 to 6000 μM with a sensitivity of 1000 μA mM −1 cm −2 and a response time of less than 2 s. The hybrid sensor also exhibited cycling durability and short-term shelf stability under the tested protocol, and strong selectivity toward glucose, with minimal interference from uric acid, ascorbic acid, dopamine, major cations, and various structurally related sugars under the tested 10:1 glucose-to-interferent screening conditions. These results support U-Ni-A/rGO as a potential platform for nonenzymatic glucose sensing.