Answers are grounded in this paper and cite their sources.
Hard X-ray micro-tomography is a powerful tool to reveal the internal structure of thick objects in a non-destructive manner. For synchrotron applications and practical lab works with students, a 2D detector based on a thin YAG:Ce scintillator optically coupled to a new generation high frame rate (100 fps) low noise sCMOS camera has been developed and characterized in depth on the METROLOGIE and PSICH'E beamlines of the SOLEIL synchrotron. The Detector gain, Modulation Transfer Function, Noise Power Spectrum and Detective Quantum Efficiency have been measured and compared with analytical model. A tomography reconstruction was performed on small insects to demonstrate the performance of this X-ray imaging detector.
No takes yet. Share an insight, caveat, or question.
Desjardins et al. (2014) studied this question.