Key result
Gallium-based liquid metal CAROL ablation achieves effective target temperatures in ~89% of porcine lung models.
Why the study?
Radiofrequency ablation for lung nodules requires exact delivery of rigid electrodes, often causing complications or suboptimal results, motivating testing of flexible conforming electrodes.
Does CAROL ablation using a liquid metal conforming electrode improve ablation efficacy in a porcine lung model?
Does CAROL ablation using a liquid metal conforming electrode improve ablation efficacy in a porcine lung model?
CAROL ablation using a liquid metal conforming electrode is effective and safe in porcine lungs, offering a potential novel approach for lung nodule ablation.
Flexible electrodes feasible for bronchoscopic RFA in porcine lungs; leaves open safety, efficacy, and translation to human nodules.
BACKGROUND: Radiofrequency ablation (RFA) is one of the treatment options for lung nodules. However, the need for exact delivery of the rigid metal electrode into the center of the target mass often leads to complications or suboptimal results. To overcome these limitations, a concept of conforming electrodes using a flexible material has been tested in this study. METHODS: A bronchoscopy-guided RFA (CAROL) under a temperature-controlled mode was tested in in-vivo and ex-vivo porcine lungs. Gallium-based liquid metal was used for turning the bronchial tree into temporary RF electrodes. A customized bronchoscopy-guided balloon-tipped guiding catheter (CAROL catheter) was used to make the procedure feasible under fluoroscopy imaging guidance. The computer simulation was also performed to gain further insight into the ablation results. Safety was also assessed including the liquid metal remaining in the body. RESULTS: The bronchial electrode injected from the CAROL catheter was able to turn the target site bronchial air pipe into a temporally multi-tined RF electrode. The mean volume of Gallium for each effective CAROL was 0.46 ± 0.47 ml. The ablation results showed highly efficacious and consistent results, especially in the peripheral lung. Most bronchial electrodes were also retrieved by either bronchoscopic suction immediately after the procedure or by natural expectoration thereafter. The liquid metal used in these experiments did not have any significant safety issues. Computer simulation also supports these results. CONCLUSION: The CAROL ablation was very effective and safe in porcine lungs showing encouraging potential to overcome the conventional approaches.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2023) studied Lung nodule (porcine model) (n=47). CAROL ablation (bronchoscopy-guided RFA with E-GaIn liquid metal) vs. Saline-assisted linear RF electrode (in subset) was evaluated on Effective CAROL ablation (central bronchial electrode temperature ≥ 60 °C). CAROL ablation using a gallium-based liquid metal to create a conforming bronchial electrode achieved effective target temperatures (≥60 °C) in 89.3% of porcine lung ablations.