Key result
Fibrous jaw clamps prevent serious intimal injury to skeletonized ITAs versus metal clamps.
Why the study?
The susceptibility of the ultrasonically skeletonized internal thoracic artery to intimal injury from vessel clamps was unclear.
Does a fibrous jaw clamp prevent intimal injury compared to a metal clamp in ultrasonically skeletonized internal thoracic arteries during CABG?
Does a fibrous jaw clamp prevent intimal injury compared to a metal clamp in ultrasonically skeletonized internal thoracic arteries during CABG?
Fibrous jaw clamps are recommended over metal clamps for temporary clamping of ultrasonically skeletonized internal thoracic arteries to prevent intimal injury during CABG.
Answers are grounded in this paper and cite their sources.
Fibrous jaw clamps may limit intimal injury during ITA clamping; leaves open effects on graft patency and outcomes.
The skeletonized internal thoracic artery (ITA) has several advantages over a pedicled one in coronary artery bypass grafting. A skeletonized ITA, which lacks surrounding tissue, thus seems more susceptible to the mechanical force exerted by a vessel clamp than the pedicled ITA. The purpose of this study was to assess the detrimental effect of vessel clamps on the intimal integrity of the ultrasonically skeletonized ITA. We skeletonized twelve ITAs with an ultrasonic scalpel in patients who underwent coronary artery bypass grafting, and thereafter two types of clamp, namely a metal clamp and a fibrous jaw clamp, were applied to the terminal portion of the ITA for 30 min. The intimal integrity of the ITAs was morphologically assessed using scanning electron microscopy. A metal clamp can cause serious intimal injury which disrupts the internal elastic lamina, and thus should be avoided for the temporary clamping of the skeletonized ITA. A fibrous jay clamp, however, hardly ever causes intimal injury, and its clinical use for the temporary clamping of the ultrasonically skeletonized ITA is therefore recommended. Vessel clamps can cause intimal injury of the ultrasonically skeletonized ITA, and the degree of the injury depends on the type of the clamp used.
No takes yet. Share an insight, caveat, or question.
Yoshikai et al. (2007) studied Coronary artery bypass grafting (n=12). Metal clamp vs. Fibrous jaw clamp was evaluated on Intimal integrity assessed using scanning electron microscopy. A metal clamp caused serious intimal injury to ultrasonically skeletonized internal thoracic arteries, whereas a fibrous jaw clamp hardly ever caused injury.
Synapse has enriched 10 related papers. Consider them for comparative context: