Proof-of-concept study demonstrates highly reliable hand pinch force and rotation measurement using an instrumented key rig, highlighting its utility for clinical rehabilitation monitoring.
Objective quantification of hand functions using sensorized devices has gained recognition lately in clinical practice. Traditional clinical assessment scales lack precision and quantifiability, prompting the exploration of new methodologies. The following work focuses on the design, development, and testing of a sensorized key rig that can be used to perform quantifiable metrics of hand functions. Leveraging advancements in sensor technology, the study developed an instrumented key rig equipped with an off-center load cell for accurate pinch force measurement. In addition, a magnetic rotary encoder was added to track the key-turning movement. Calibration and testing procedures ensured the rig’s reliability and repeatability. Results demonstrated consistent and precise measurements with the reliability intraclass correlation coefficient (ICC) ofp \,\, =0.99996, highlighting the key-rig device potential for clinical motor function assessment and rehabilitation monitoring. Further research could explore its comparative effectiveness and clinical applications with stroke survivors.
No takes yet. Share an insight, caveat, or question.
Soh et al. (2024) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: