Key result
Leadless pacemaker implantation can result in PVC-induced polymorphic ventricular tachycardia secondary to implant position.
Why the study?
Leadless pacemaker implantation results in unique complications distinct from traditional device implantation, including ventricular arrhythmias potentially related to device implant position.
Case Report
Ventricular arrhythmias following leadless pacemaker implantation may be secondary to the implant position and should be recognized as a unique complication.
Answers are grounded in this paper and cite their sources.
May warrant position reassessment in leadless pacemaker patients with new ventricular arrhythmias; case report leaves open incidence and prevention strategies.
Key Teaching Points•Leadless pacemaker implantation will result in unique complications distinct from traditional device implantation. Ventricular arrhythmias arising following implantation of a leadless pacing system should be recognized as potentially secondary to the implant position of the device.•Leadless pacing systems have no ability to store ventricular high-rate electrograms. Episodes of ventricular high rates should raise concern for ventricular arrhythmias in these patients.•Extraction of leadless devices is safe and effective. An experienced extraction team should be available at centers implanting leadless pacing devices. •Leadless pacemaker implantation will result in unique complications distinct from traditional device implantation. Ventricular arrhythmias arising following implantation of a leadless pacing system should be recognized as potentially secondary to the implant position of the device.•Leadless pacing systems have no ability to store ventricular high-rate electrograms. Episodes of ventricular high rates should raise concern for ventricular arrhythmias in these patients.•Extraction of leadless devices is safe and effective. An experienced extraction team should be available at centers implanting leadless pacing devices.
No takes yet. Share an insight, caveat, or question.
Amin et al. (2018) conducted a case report in Premature ventricular contraction-induced polymorphic ventricular tachycardia. Leadless pacemaker implantation was evaluated. Leadless pacemaker implantation can result in unique complications such as premature ventricular contraction-induced polymorphic ventricular tachycardia secondary to the implant position.