Key result
Developing pneumothorax during active can ICD implantation can impede current flow and prevent effective defibrillation.
Why the study?
Several unsuccessful defibrillations during active can ICD implantation with subclavian vein puncture may be related to pneumothorax affecting defibrillation current flow.
Case Report (n=1)
In cases of unsuccessful defibrillations during active can ICD implantation involving subclavian vein puncture, clinicians should consider the possibility of a pneumothorax altering current flow.
Answers are grounded in this paper and cite their sources.
May warrant pneumothorax evaluation during failed active-can ICD defibrillation; leaves open whether targeted interventions improve outcomes.
During implantable cardioverter defibrillator (ICD) implantation of an active can ICD several defibrillations with 20 J and 34 J as well as 360 J externally were ineffective. The implant criteria were finally met with a second defibrillation lead and reversed polarity. A left-sided pneumothorax due to subclavian vein puncture was detected soon after ICD implantation. It is assumed that especially in the active can alignment the developing pneumothorax made defibrillation current flow more difficult. In case of several unsuccessful defibrillations during active can ICD implantation in which the subclavian vein was punctured, the possibility of a pneumothorax should be considered.
No takes yet. Share an insight, caveat, or question.
Schuchert et al. (1998) conducted a case report in Pneumothorax during active can ICD implantation (n=1). Active can ICD implantation was evaluated on Defibrillation success. A developing pneumothorax during active can ICD implantation can impede defibrillation current flow, leading to ineffective internal and external defibrillations.
Synapse has enriched 10 related papers. Consider them for comparative context: