Why the study?
Does pneumothorax drainage improve defibrillation thresholds in patients with ICD defibrillation failure due to pneumothorax?
Does pneumothorax drainage improve defibrillation thresholds in patients with ICD defibrillation failure due to pneumothorax?
Pneumothorax should be considered in the differential diagnosis of unexpectedly high defibrillation thresholds during ICD implantation, which can be resolved by drainage.
Answers are grounded in this paper and cite their sources.
Supports considering pneumothorax in unexplained high DFTs at ICD implant; single case leaves open broader applicability.
We describe two patients with defibrillation failure of implantable cardioverter defibrillators (ICDs) resulting from large left pneumothoraxes following subclavian vein puncture during the implantation. Following pneumothorax drainage, low defibrillation thresholds (DFTs) were attained without further manipulations. The absence of other signs and symptoms of pneumothorax and the presence of satisfactory pacing function during the procedure, resulted in a significant delay in diagnosis. Pneumothorax should be included in the differential diagnosis when unexpected high DFTs are found during ICD implantation or predischarge testing. This complication is avoidable by a different surgical approach, cephalic vein cutdown.