Key result
Underexpanded and overexpanded bioresorbable scaffolds dynamically remodel at 5 years to normalize endothelial shear stress.
Case Report (n=2)
The study provides five-year follow-up data on lumen dimensions and endothelial shear stress in underexpanded and overexpanded bioresorbable scaffolds.
Answers are grounded in this paper and cite their sources.
Device-specific ESS and lumen metrics post-scaffold implantation differ; leaves open prognostic value for clinical events.
Median ESS: 1.41 (0.69, 2.49) Pa Post implantation: In-scaffold area stenosis: 47.4% Mean lumen area: 5.58 mm 2 Mean lumen diameter: 2.64 mm Median ESS: 1.58 (1.05, 2.48) Pa Post implantation: In-scaffold area stenosis: 49.19% Mean lumen area: 6.39 mm 2 Mean lumen diameter: 2.84 mm Median ESS: 3.51 (1.53, 6.08) Pa Post implantation: In-scaffold area stenosis: 40.49%Mean lumen area: 4.01 mm 2 Mean lumen diameter: 2.25 mm Median ESS: 2.65 (1.89, 4.42) Pa
No takes yet. Share an insight, caveat, or question.
Torii et al. (2017) conducted a case report in Coronary artery disease (n=2). Absorb bioresorbable vascular scaffolds was evaluated on Endothelial shear stress (ESS) normalisation. At 5-year follow-up, both underexpanded and overexpanded bioresorbable scaffolds underwent dynamic remodelling that normalised endothelial shear stress to physiological ranges.