Key result
Transfemoral TAVR with ACURATE neo achieves ~88% device success in severe native aortic regurgitation.
Why the study?
TAVR has been proposed as an alternative to medical therapy in inoperable patients with severe native aortic valve regurgitation, but safety and efficacy data are limited.
Does transfemoral TAVR with the ACURATE neo valve provide safe and effective treatment in inoperable patients with severe native aortic valve regurgitation?
Observational (n=24)
Yes
Does transfemoral TAVR with the ACURATE neo valve provide safe and effective treatment in inoperable patients with severe native aortic valve regurgitation?
Transfemoral TAVR with the ACURATE neo valve is feasible and shows good early safety in inoperable patients with severe native aortic regurgitation, though it carries higher risks of paravalvular leak and pacemaker implantation compared to TAVR for aortic stenosis.
May support feasibility in native AR but should not change practice; leaves open randomized evaluation of ACURATE neo.
Transcatheter aortic valve replacement (TAVR) has been validated for the treatment of severe symptomatic aortic stenosis in patients at high and intermediate surgical risk. Recently, TAVR has been proposed as an alternative to medical therapy in inoperable patients with severe native aortic valve regurgitation (NAVR). This multicenter international registry sought to evaluate safety and efficacy of TAVR with the self-expandable ACURATE neo valve in a cohort of patients with NAVR. A total of 24 patients with severe NAVR treated by TAVR between September 2016 and October 2018 in 13 European centers were included. Clinical, procedural and follow up data were inserted in a dedicated database. Outcomes were codified according to Valve Academic Research Consortium-2 criteria. Mean age was 79.4 years, 58.4% were female. Mean EuroSCORE II and STS score were 5% and 3.9%, respectively. Device success was 87.5%. Moderate paravalvular leak (PVL) was found in two (8.3%) of patients, both with a perimeter oversizing index <10%. Implantation of a second device was necessary in three cases (12.5%), one for severe PVL and two for device displacement. New pacemaker implantation rate was 21.1%. At 30 days, stroke and all-cause mortality rates were 0% and 4.1%, respectively. This multicenter study suggests good feasibility and early safety of transfemoral TAVR with the self-expandable ACURATE neo device in patients with severe NAVR refused for surgery. Rates of moderate PVL, new pacemaker implantation and need for a second valve were higher than those reported for TAVR in aortic stenosis.
No takes yet. Share an insight, caveat, or question.
Purita et al. (2020) conducted an observational in severe native aortic valve regurgitation (n=24). Transcatheter aortic valve replacement with ACURATE neo valve was evaluated on Device success. Transfemoral TAVR with the ACURATE neo device in patients with severe native aortic valve regurgitation achieved 87.5% device success and 4.1% 30-day mortality.