Key result
S-ICD fails to terminate monomorphic VT despite 5 consecutive 80-J shocks in HCM.
Why the study?
Subcutaneous implantable cardioverter-defibrillators are generally safe and effective, but rare cases of ineffective shock therapy in hypertrophic cardiomyopathy patients have been reported.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
This case report documents a rare failure of an S-ICD to terminate monomorphic ventricular tachycardia with 80-J shocks in a patient with hypertrophic cardiomyopathy.
We report the case of an 18-year-old hypertrophic cardiomyopathy female with a subcutaneous implantable cardioverter-defibrillator (S-ICD) for primary prevention who presented with monomorphic ventricular tachycardia and five ineffective 80-J shocks.We have examined the patient case notes, imaging and electrophysiology study data, and traces obtained from S-ICD device interrogation.This is a rare example of device failure in an otherwise safe and effective system as illustrated by EFFORTLESS and IDE data.
No takes yet. Share an insight, caveat, or question.
Bunny Saberwal (2016) conducted a case report in Hypertrophic cardiomyopathy (n=1). Subcutaneous implantable cardioverter-defibrillator (S-ICD) was evaluated on Efficacy of S-ICD shock therapy for monomorphic ventricular tachycardia. A subcutaneous implantable cardioverter-defibrillator delivered five ineffective 80-J shocks for monomorphic ventricular tachycardia in an 18-year-old female with hypertrophic cardiomyopathy.