Key result
Advanced 3D imaging enables successful transvenous LV lead placement after laser extraction in persistent LSVC.
Why the study?
Laser lead extraction of implantable cardioverter defibrillator leads via a persistent left superior vena cava had not been previously reported, and advanced imaging techniques for transvenous LV lead implantation in this context were not well described.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
This is the first reported case of successful laser extraction of an ICD lead via a persistent left SVC, highlighting the value of advanced 3D imaging overlay for complex lead implantations.
A 60-year-old man underwent attempted cardiac resynchronization with defibrillator therapy utilizing a left-sided approach. A persistent left-sided superior vena cava (SVC) was identified and a dual-coil defibrillator lead (Sprint Fidelis; Medtronic) and atrial lead were deployed via this route (Figure 1A). No left ventricular (LV) branches of the coronary sinus (CS) were identified and so an LV pacing lead was implanted surgically onto the anterior aspect of the LV via a mini-thoracomtomy. The patient presented 5 years later with inappropriate shocks secondary to suspected fracture of the defibrillator lead. Laser lead extraction of the defibrillator lead via the left SVC was performed successfully. A contrast-enhanced cardiac computed tomography scan was performed as a prelude to extraction and a three-dimensional (3D) segmented anatomy of the cardiac chambers was derived using fully automated prototype segmentation software (Philips Healthcare). Further manual segmentation was performed to identify the main CS and branches. The 3D model was overlaid onto real-time fluoroscopy using prototype software (Philips Healthcare). This enabled successful implantation of a transvenous LV pacing lead into a more conventional posterolateral position (Figure 1C). To our knowledge, this is the first report of successful laser lead extraction of an implantable cardioverter defibrillator lead via a persistent left SVC. Our case also demonstrates the value of advanced imaging techniques in difficult cases of transvenous LV lead implantation. The full-length version of this report can be viewed at: http://www.escardio.org/communities/EHRA/publications/ep-case-reports/Documents/laser-extraction.pdf
No takes yet. Share an insight, caveat, or question.
Sohal et al. (2013) conducted a case report in Suspected defibrillator lead fracture with persistent left superior vena cava (n=1). Laser lead extraction and 3D segmented anatomy overlay was evaluated on Successful laser lead extraction and transvenous LV pacing lead implantation. Laser lead extraction of a defibrillator lead via a persistent left superior vena cava was performed successfully, with advanced 3D imaging enabling transvenous LV lead implantation.