Key result
GORE CARDIOFORM ASD occluder achieves ~95% implantation success in pediatric and adult patients.
Why the study?
The learning curve and early outcomes of using the GORE CARDIOFORM atrial septal defect occluder for ASD closure in paediatric and adult patients were not well understood.
Does the GORE CARDIOFORM ASD Occluder provide successful and safe atrial septal defect closure in paediatric and adult patients?
Observational (n=135)
Yes
Does the GORE CARDIOFORM ASD Occluder provide successful and safe atrial septal defect closure in paediatric and adult patients?
The GORE CARDIOFORM ASD Occluder demonstrates a high implantation success rate (95%) for ASD closure, though new onset arrhythmia occurred in 12% of patients and was associated with larger device sizes.
May inform conformable device selection in pediatric ASD closure; leaves open prospective validation of learning curve effects.
The GORE CARDIOFORM atrial septal defect (ASD) Occluder (GCA) is composed of a platinum-filled nitinol wire frame covered with expanded polytetrafluoroethylene, making it softer and more conformable compared with nitinol mesh devices. After the ASSURED clinical study confirmed the efficacy and safety of the device, it received U.S. Food and Drug Administration approval and a European conformity mark. Our aim was to understand the learning curve implicated in using the GCA for ASD closure in paediatric and adult patients as well as to study the early outcomes. To this end, a review of ASD device closures with GCA in 4 UK centres was conducted between January 2020 and January 2023. Implantation success was the primary outcome; the secondary outcomes were serious adverse events, including new onset arrhythmia. In all, 135 patients were included, and 128 (95%) had successful ASD device closure with GCA. The median patient age was 49 years, the median defect size was 18 mm, and the median device size was 37 mm. The median follow-up time was 6 months (interquartile range 1-14). One device embolisation occurred, and 15 patients (12% of GCA implantations) developed new onset arrhythmia - this was not related to patient age, defect diameter or device oversizing but was positively associated with device size. With growing experience using GCA, the device can be applied to a wide variety of ASD sizes and morphologies. Given the number of successful implantations with an absence of aortic erosion, as well as the ability to perforate through the device should procedures be required in the left atrium, the GCA device is an important addition for interventionists who close atrial septal defects.
No takes yet. Share an insight, caveat, or question.
Hribernik et al. (2023) conducted an observational in Atrial septal defect (ASD) (n=135). GORE CARDIOFORM atrial septal defect (ASD) Occluder (GCA) was evaluated on Implantation success. The GORE CARDIOFORM atrial septal defect occluder achieved a 95% implantation success rate in paediatric and adult patients, with 12% developing new onset arrhythmia.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: