Key result
Rescue deployment of the GORE cardioform atrial septal defect occluder successfully closed a large ASD without recurrence of complete atrioventricular block in a 7-year-old child.
Why the study?
Atrioventricular block is a life-threatening complication of transcatheter ASD closure related to device stiffness and defect anatomy, making the new GORE cardioform ASD occluder an appealing technical alternative.
Case Report (n=1)
No
The GORE cardioform ASD occluder may serve as a safe rescue alternative for ASD closure in patients who develop atrioventricular block with stiffer self-centering devices.
Answers are grounded in this paper and cite their sources.
May offer rescue option for pediatric ASD with AV block; leaves open need for larger studies before wider use.
Atrioventricular block (AVB) is an infrequent but life-threatening complication of transcatheter closure of atrial septal defect (ASD), accounting for 0.1%–6.2% of cases in large series. It has been related to unfavorable defect anatomy as well as size and intrinsic stiffness of the occluding device. In this setting, the new GORE ® cardioform ASD occluder (GCA) device could be an appealing technical advance in ASD treatment. We report a case of complete AVB after ASD closure with an Amplatzer septal occluding (Abbott, Plymouth MN, USA) device successfully treated by its percutaneous retrieval and “rescue” deployment of GCA device few months later.
No takes yet. Share an insight, caveat, or question.
Pizzuto et al. (2022) conducted a case report in Atrial septal defect (n=1). GORE cardioform ASD occluder was evaluated on Successful ASD closure and resolution of atrioventricular block. Rescue deployment of the GORE cardioform atrial septal defect occluder successfully closed a large ASD without recurrence of complete atrioventricular block in a 7-year-old child.
Synapse has enriched 10 related papers. Consider them for comparative context: