Key result
Pediatric RFCA yields maximum skin radiation doses up to ~2.4 Gy.
Why the study?
Because RF catheter ablation procedures may be lengthy and are commonly performed in young patients, concern has arisen about radiation dose in this group of patients.
Observational (n=9)
Pediatric patients undergoing RF catheter ablation can receive significant radiation skin doses, highlighting the need to measure dose rates for individual equipment to establish procedural time limits.
Answers are grounded in this paper and cite their sources.
Skin dose variability in pediatric ablation warrants monitoring caution; leaves open prospective evaluation of long-term risks.
Because RF catheter ablation procedures may be lengthy and are commonly performed in young patients, concern has arisen about radiation dose in this group of patients. This article investigates radiation doses in pediatric patients undergoing RF catheter ablation. Standard fluoroscopic equipment used for diagnostic electrophysiological catheterization studies is technologically capable of dose rates as high as 90 milligray (mGy) per minute to skin and adjacent lung and 260 mGy/min to vertebral bone. Dose rates of this magnitude when used for extended periods of time have been known to cause erythema, pneumonitis, and retardation of bone growth. We measured skin dose rates of nine pediatric patients undergoing RF catheter ablation for tachycardia and calculated doses to the skin using standard dosimetric methods. Fluoroscopic techniques and equipment were studied using a patient simulating phantom. Overlap of fluoroscopic fields was checked using radiotherapy portal verification film, and regions in which doses overlapped from multiple angle exposures were verified using a treatment planning computer. Patient skin dose rates ranged from 6.2-49 mGy/min for patients ranging in age from 2-20 years. Maximum skin doses ranged from 0.09-2.35 Gy. Our data demonstrate the need to directly measure dose rates for individual fluoroscopic equipment and procedural techniques in order to determine whether limitations need to be set for procedural times.
No takes yet. Share an insight, caveat, or question.
Geise et al. (1996) conducted an observational in Tachycardia (n=9). Radiofrequency catheter ablation was evaluated on Skin dose rates and maximum skin doses. Radiofrequency catheter ablation in pediatric patients resulted in skin dose rates ranging from 6.2 to 49 mGy/min and maximum skin doses ranging from 0.09 to 2.35 Gy.