Key result
Electrocautery coagulation mode increases lead damage to 48% compared with 2% for cutting mode.
Why the study?
Electrocautery can damage transvenous cardiac device leads, but the impact of different electrocautery techniques and insulated blades on lead damage is unclear.
Does the choice of electrocautery technique and blade type affect the incidence of damage to transvenous cardiac device leads?
Does the choice of electrocautery technique and blade type affect the incidence of damage to transvenous cardiac device leads?
Absolute Event Rate: 48% vs 2%
p-value: p=<0.0001
Using cutting mode, flat blade orientation, and the PhotonBlade minimizes electrocautery-induced damage to transvenous cardiac device leads.
Cautery technique and blade type were associated with varying in vitro lead damage; leaves open whether findings guide clinical practice.
BACKGROUND: Electrocautery (cautery) can damage transvenous cardiac device leads. The purpose of this study was to compare lead damage from an insulated cautery blade when used with several different techniques that included coagulation (COAG) versus cutting (CUT) mode, perpendicular active edge (active) versus parallel flat blade (flat) orientation (phase 1), and using one commercially available blade (PhotonBlade) versus another (PlasmaBlade) (phase 2). METHODS: In phase 1, lesions were delivered using combinations of: (1) COAG and CUT; (2) active and flat orientation; and (3) polyurethane, silicone, and copolymer insulation. In phase 2, lesions were delivered using combinations of: (1) PlasmaBlade and PhotonBlade, (2) four power output levels, and (3) eight different lead models. Lead damage was scored on an ordinal scale of 0 to 4. RESULTS: Phase 1: more leads were damaged using COAG than CUT (48% vs 2%, P < 0.0001). When using COAG, 74% of lesions using active orientation had damage versus 22% of lesions using flat orientation (P = 0.0002). COAG lesions to copolymer (61%) and polyurethane (68%) leads had greater damage than silicone (17%) (P = 0.006 and P = 0.003, respectively). Phase 2: 75% of treatments using PlasmaBlade had damage versus 40% of treatments with PhotonBlade (P < 0.0001). Higher power resulted in more damage. At the commonly used setting of CUT 20 W, damage occurred in 39% of treatments using PlasmaBlade versus 13% using PhotonBlade (P = 0.0006). CONCLUSIONS: COAG resulted in more damage than CUT; this effect was greatest with the active edge, and with polyurethane or copolymer insulation. PhotonBlade was associated with less damage to leads than PlasmaBlade.
No takes yet. Share an insight, caveat, or question.
Wasserlauf et al. (2018) studied Transvenous cardiac device leads. Coagulation (COAG) mode vs. Cutting (CUT) mode was evaluated on Lead damage (p=<0.0001). Coagulation mode damaged more leads than cutting mode (48% vs 2%, P<0.0001), and PlasmaBlade caused more damage than PhotonBlade (75% vs 40%, P<0.0001).