A serious impediment to precision metrology with scanning probe microscopes is the unreproducible, nonlinear behavior of the piezo ceramic actuators. We have developed a simple solution to the problem by measuring the position of the scanning head with capacitors. The circuit monitoring the gap between the plates has a linear response with slope 12 μm/V and noise of 0.1 nm/(Hz)1/2. The linearity of the system is verified by comparing the capacitor output with a scan of a grating that is periodic in two dimensions. We analyze the errors associated with the technique and show how to reduce them to acceptable levels.