Key result
The Engström metabolic computer measured oxygen consumption and CO2 production with mean errors generally less than 1% compared with predicted values, though errors in VO2 were erratic at FIO2 ≥0.7.
The Engström metabolic computer provides accurate measurements of VO2 and CO2 production at lower FIO2 levels, but becomes erratic at FIO2 ≥ 0.7.
This is the authors' abstract. We don't add key points for this paper.
Laboratory evaluation and use of the Engström metabolic computer (EMC) in a clinical setting are described. The accuracy of the EMC was tested using an inert gas dilution technique. Mean errors in oxygen consumption (VO2) and CO2 production compared with predicted values were less than 2% of predicted, and generally less than 1%, but with an SD of less than or equal to 5.1%. At an FIO2 of greater than or equal to 0.7, the errors in VO2 were erratic and generally greater than 15%. The apparatus has the facility to assume an RQ in the calculation of VO2 and the errors in measured VO2 were less than 1% up to an FIO2 of 0.8.
Answers are grounded in this paper and cite their sources.
Caution against use at FIO2 ≥0.7; leaves open prospective clinical validation of the Engström computer.
No takes yet. Share an insight, caveat, or question.
Regan et al. (1990) studied this question. Engström metabolic computer vs. Inert gas dilution technique (predicted values) was evaluated on Mean errors in oxygen consumption (VO2) and CO2 production. The Engström metabolic computer measured oxygen consumption and CO2 production with mean errors generally less than 1% compared with predicted values, though errors in VO2 were erratic at FIO2 ≥0.7.