Key result
Noninvasive platelet monitoring devices fail to adequately assess concentrate quality compared to standard methods.
Why the study?
The accuracy of the Platelet Monitoring System and NAPSAC instruments in noninvasively assessing platelet concentrate quality during storage was uncertain.
Do the PMS and NAPSAC instruments accurately assess the quality of platelet concentrates compared to standard methods?
Observational (n=134)
Do the PMS and NAPSAC instruments accurately assess the quality of platelet concentrates compared to standard methods?
Neither the PMS nor the NAPSAC instrument adequately assesses the quality of PL732 platelet concentrates noninvasively.
Answers are grounded in this paper and cite their sources.
These instruments should not replace standard platelet quality testing; leaves open development of accurate noninvasive monitors.
The Platelet Monitoring System (PMS) and the Non-invasive Assessment of Platelet Shape and Concentration (NAPSAC) instruments which relate light scattering characteristics of platelet concentrates (PC) to platelet concentration and shape, were evaluated to determine their accuracy in assessing platelet quality during storage from 1 to 7 days. The results were correlated with platelet concentration, % discs and pH on 121 PC stored in PL732 containers. NAPSAC output is in the form of platelet concentration and % discs. When NAPSAC and standard method values were compared, correlation coefficients (r) did not exceed 0.76 for counts and 0.62 for % discs. PMS output is in the form of lights with red indicating poor quality and green or amber indicating acceptable quality. Sensitivity of the PMS instrument did not exceed 83% and specificity did not exceed 63%. Mean platelet number, % discs and pH were comparable for units triggering red versus green or amber lights. In a separate study, 13 PL732 PC stored 5 days and transfused autologously were evaluated on the PMS. Three red light units exhibited recovery and survival times similar to those observed with PC triggering green/amber lights. These data indicate that neither instrument adequately assesses the quality of PL732 PC.
No takes yet. Share an insight, caveat, or question.
George et al. (1989) conducted an observational in Platelet concentrate quality assessment (n=134). Platelet Monitoring System (PMS) and NAPSAC instruments vs. Standard method values (platelet concentration, % discs, and pH) was evaluated on Accuracy in assessing platelet quality (correlation for NAPSAC, sensitivity and specificity for PMS) (r ≤ 0.76 for counts, r ≤ 0.62 for % discs; sensitivity ≤ 83%, specificity ≤ 63%). The Platelet Monitoring System and NAPSAC instruments did not adequately assess platelet concentrate quality, demonstrating poor correlation and low sensitivity versus standard methods.